According to DIN EN 60751 it is necessary to publish the response sensitivity ( half-life tO,5) of sensor resistances and thermometers. The answer signal of the temperature sensor is noted on a temperature jump. The measurement takes place in the air flow of 1m/s and in flowing water of 0,4m/s.
These half-lives amount to in water for
These large spreads indicate that these values are strongly coupled to the design of the sensor resistance.
With a measurement insert, e.g. type according to DIN 43762, this half-life lies between 6 and 9 seconds.
With a thermo sensor with measurement insert and protection fitting, e.g. according to DIN 43772, between 36 and 40 seconds.
These changes and spreads depend on the following factors: